ChevronTM Start-up Procedure - Electrical Test Procedure

Getting Started

NOTE:

  • Read the entire start-up procedure before applying any voltages.
  • Refer to Diagram 1 - Typical Wiring Diagrams - for each detection mode.
  • The suggested bias voltage for a Resistive Anode Encoder (RAE) is 300 volts.     

CAUTION:

  • Do not exceed 1000V per Microchannel Plate (MCP) for 40:1 
    aspect ratio MCPS. 1200V/MCP for 60:1 aspect ratio MCPS.
  • When installing flange mounted detectors gradually tighten the bolts in a star pattern (DO NOT exceed 20 foot-pounds per bolt). Failure to do so could cause the fiberoptic to crack.

RECOMMENDATION:

  • For optimal lifetime, operate the detector at the minimum voltage necessary to obtain a useable signal.
  • Do not operate the phosphor screen at a higher than recommended potential.

 

Procedure

Make all connections to the assembly. Check all electrical connections for possible shorted or open circuits.

Pump down to 2x10-6 torr and hold for at least 15 hours.

 

Voltage Application:

Electron/Negative Ion/UV Photon Detection: (for a metal anode or Resistive Anode Encoder, skip to next section)

Phosphor Screen

Ground the input of the assembly (Vi). Apply voltage to the phosphor screen (Va) in +250V, 1 minute increments. Stop at +1.0 kV.

Apply voltage to the output of the assembly (Vo) in +100V, 2 minute increments. Stop at +1.0 kV.

Increase the voltage to Va in +100V, 5 minute increments to +3.0 kV. Wait 5 minutes.

Increase the voltage to Va in +100V, 10 minute increments to +4.0 kV. Wait 5 minutes.      

Simultaneously increase the voltage to Va and Vo in +100V, 10 minute increments to +4.5 kV at Va and +1.5 kV at Vo.

For screens requiring a 5.0 kV potential - Increase the voltage to Va in +100V, 10 minute increments to +5.5 kV. Wait 10 minutes.

For screens requiring a 5.0 kV potential - Increase the voltage to Va in +50V, 10 minute increments to +6.5 kV. Wait 10 minutes.

Simultaneously increase the voltage to Va and Vo in +50V, 10 minute increments to +2.0 kV at Vo.

When through using the detector, turn off the voltage to Va. When the voltage drops below +2.0 kV, turn off the voltage to Vo. 

Metal Anode/Resistive Anode Encoder

Ground the input of the assembly (Vi). Apply the specified anode bias to Va.

Increase the voltage to both Va and Vo in +100V, 2 minute increments by +1.0 kV at Va and to +1.0 kV at Vo. Wait 5 minutes.

Increase the voltage at Vo and Va in +100V, 5 minute increments to +1.5 kV at Vo. Wait 10 minutes.

Increase the voltage at Vo and Va in +50V, 5 minute increments to +2.0 kV at Vo.

When through using the detector, turn off the voltages to Vo and Va.      

Positive Ion/UV Photon Detection (for a metal anode or Resistive Anode Encoder, skip to next section).

Phosphor Screen

Ground the output of the assembly (Vo). Apply voltage to the phosphor sheen (Va) in +250V, 1 minute increments. Stop at +1.0 kV.

Apply voltage to the input of the assembly (Vi) in -100V, 2 minute increments. Stop at -1.0 kV.

Increase the voltage to Va in +100V, 5 minute increments to +2.0 kV. Wait 5 minutes.

Increase the voltage to Va in +100V, 10 minute increments to +3.0 kV. Wait 5 minutes.

Adjust the voltage to Vi in -100V, 10 minute increments to -1.5 kV.      

For screens requiring a 5.0 kV potential - Increase the voltage to Va in +100V, 10 minute increments to +4.0 kV. Wait 10 minutes.

For screens requiring a 5.0 kV potential - Increase the voltage to Va in +50V, 10 minute increments to +5.0 kV. Wait 10 minutes.

Adjust the voltage to Vi in -50V, 10 minute increments to -2.0 kV.

When through using the detector, turn off the voltages to the Vi and Va.      

Metal Anode/Resistive Anode Encoder

Ground the output of the assembly (Vo). Apply the specified anode bias to Va.

Apply voltage to Vi in -100V, 2 minute increments. Stop at -1.0 kV. Wait 2 minutes.

Adjust the voltage at Vi in -100V, 5 minute increments to -1.5 kV. Wait 5 minutes.

Adjust the voltage at Vi in -50V, 10 minute increments to -2.0 kV.

When through using the detector, turn off the voltages to Vi and Va.      

 